Image

PHILIPS BV 300

The PHILIPS BV 300 Specialty Packages Available
Extended processing
• Vascular functions
• Extended vascular functions.
Generator
• 2.1 kW full-wave.
• Up to 110 kVp.
• Up to 20 mA for radiographic film exposures.
• Continuous fluoro boost mode up to 7.3 mA
Full power from outlet:
• 50 Hz or 60 Hz.
• 100V-15A to 240V-10A.
• X-Ray Tube
Fixed anode X-ray tube.
• Focal spots: 0.6 mm and 1.4 mm.
• Anode heat capacity: 50,100 HU.
• Anode cooling rate: 20,200 HU/min.
• Housing heat capacity: 1,200,000 HU.
• Anode angle: 12°.
• Collimator
• Iris collimation with dual opposing lead shutters allowing elliptical, round and rectangular collimation.
• Minimum beam size < 5 X 5 cm2.
• Iris and shutters are continuously variable.
• All functions remotely controlled from the C-Arm control panel.
• Fluoroscopy Mode
Focal spot: 0.6 mm.
• kVp range: 40 – 110 kVp.
• mA range: 0.1 – 3.1 mA Low Dose fluoro mode.
• mA range: 0.2 – 7.3 mA continuous High Definition fluoro boost mode.
• Auto and manual modes.
• Continuous or Pulsed operation.
• APF varies mA, kVp and camera gain.
• User specific APF tables.
• Pulsed Fluoroscopy Mode (optional)

 